TKA 2 Microplate
Laboratories that perform few analyses can afford to equip themselves with one of these "walk-away" analyzers, thus avoiding the use of several machines. This helps in reducing costs, saving room, and protecting against human error. The programs allow the creation of profiles containing up to 8 assays to be carried out in a single session. It is also possible to carry out different assays on the same microplate, provided that such assays have the same temperature and incubation schedule. Like all TKA Analyzers the software of these units allow the storage of Standard Curves that can be retrieved later and used to adjust one or two calibrators, reducing analysis cost. Reagent containers are included in the kit.
TKA 2, 3, 6 and 8 Microplate Automated EIA Analyzers

Technical Specifications

TKA 6 and 8 Microplate
These are two top models of the new line and are particularly adequate to perform routine screening tests in laboratories having huge workloads and in blood banks, especially due to the use of eight needles. A whole microplate is processed in approximately 5 minutes (100 μl in individual quotas). Each incubator can freely programmed at an individual temperature. Futhermore, the instrument can be provided with the “STAT function" option that allows samples and new analysis, if so required, during the session.
2 Plates
6 Plates
Internally and externally coated in order to minimize the carry-over
8
1
Independent for each needle
10 pi
10-6 con sufficient needle washing
Option
8
8
12
Standard Washable Needles
Number of Needles
Liquid Level Sensor
Sample Minimum Volume
Pipetting Precision and Accuracy
Carry-Over of Washable Needles
Conductive Disposable Tips
Pump for Needle Washing
Maximum Number of Assays for Profile
Maximum Number of Microplates
Number of Independent Incubators
Type
Closed and darkened
Temperature
It can be programmed from 2°C AT/ 50°C, at 0.1°C step
Temperature Accuracy
±2°C
Incubation Time
From 1 minute to 24 hours, to be programmed
Microplate Shaking
At intervals, to be programmed
Constant Microplate Shaking
Option
Wavelength Range:
400/750nm (320/800nm on request)
Number of Standard Filters:
Standard Filters (nm):
Max. Number of Filter Positions:
Up to 6, automatically selected
Accuracy:
<1% at 450nm 1.0 OD
Linearity:
±1% at 450nm
Reproducability:
>±0.3% at 450nm, 1.0 OD (using neutral filter)
Range:
From 0 to 3 OD
Reading Speed:
< 5 seconds for single wavelength measurement
Kind of Reading:
Single/dual wavelength measurements
There are no items in this list
Manifold
8 distribution channels and 8 aspiration channels
Number of Washing Buffers
Dispensing Speed
To be programmed
Soak Time
To be programmed
Strip position
Automatic control of strip absence/presence
850 Watts
750 Watts
650 Watts
600 Watts
110/230 V, 50/60 Hz
15ºC - 35ºC (59ºF - 95ºF)
10% - 85% at 30ºC (85ºF) or below
1350 mm
1670 mm
2070 mm
1670 mm
790 mm
810 mm
Hard disk, floppy, CD, keyboard, color monitor, and mouse included
RS 232 and parallel
Conductive disposable tips, bar code reader
Miscellaneous
Reader
Incubators
Washer
Configuration
TKA 3 Microplate
A high performance and profitability line, due to simultaneous action of 4 needles that distribute 96 samples in the microplate within a very short period. In order to avoid contamination each needle is fitted with its own membrane pump for internal washing purpose. The operator can freely choose up to a maximum of twelve assays to be included in the Profile to carry out. This unit is an adequate solution for medium-size laboratories.

3 Plates
4
8 Plates
1
Better than 1°/0 CV at 10°/0 syringe volume, based on gravimetric evaluation
1
1
4
8
12
2
3
6
8
3
3
4
4
2
3
6
8